Wholesale China Online Dissolved Oxygen Analyzer Quotes Manufacturer - DOG-208F Industrial Dissolved Oxygen Sensor – BOQU Detail:

DOG-208F Dissolved Oxygen Electrode applicable for Polarography Principle

With platinum (Pt) as cathode and Ag / AgCl as anode
The electrolyte is 0.1 M potassium chloride (KCI).
The silicone rubber permeable membrane imported from U.S. serves as the permeable
membrane.
It has silicone rubber and steel gauze.
It is featured by collision resistance, corrosion resistance, high.temperature resistance, shape
retention and other performances
Measuring range: 0-100ug/L 0-20mg/L
Electrode material: 316L stainless steel
Temperature compensation resistor: 2.252K 22K Ptl00 Ptl000 etc
Sensor life: >3 years
Cable length: 5m (double shielded)
Detection lower Limit: 0.1ug/L(ppb)(20℃)
Measurement upper limit: 20mg/l(ppm)
Response time: ≤3min(90%, 20℃)
Polarization time: >8h
Minimum flow rate: 5cm/s; 515 L/h
Drift: <3%/month
Measurement error: <±1 ppb
Air current: 50-80nA   Note: Maximum current 20-25 uA
Polarization voltage: 0.7V
Zero Oxygen: <5ppb(60min)
Calibration intervals: >60 days
Measured water temperature:0~60℃
Measuring range: 0-100ug/L 0-20mg/L
Electrode material: 316L stainless steel
Temperature compensation resistor: 2.252K 22K Ptl00 Ptl000 etc
Sensor life: >3 years
Cable length: 5m (double shielded)
Detection lower Limit: 0.1ug/L(ppb)(20℃)
Measurement upper limit: 20mg/l(ppm)
Response time: ≤3min(90%, 20℃)
Polarization time: >8h
Minimum flow rate: 5cm/s; 515 L/h
Drift: <3%/month
Measurement error: <±1 ppb
Air current: 50-80nA   Note: Maximum current 20-25 uA
Polarization voltage: 0.7V
Zero Oxygen: <5ppb(60min)
Calibration intervals: >60 days
Measured water temperature:0~60℃

 

Applied to thermal power plants, power plant desalted water,.

boiler feed water etc places of trace oxygen content.


Product detail pictures:

Wholesale China Online Dissolved Oxygen Analyzer Quotes Manufacturer - DOG-208F Industrial Dissolved Oxygen Sensor  – BOQU detail pictures


Related Product Guide:

Our advantages are reduced prices,dynamic product sales workforce,specialized QC,solid factories,superior quality services for Wholesale China Online Dissolved Oxygen Analyzer Quotes Manufacturer - DOG-208F Industrial Dissolved Oxygen Sensor – BOQU , The product will supply to all over the world, such as: Sudan, Bulgaria, Swiss, we are fully determined to control the whole supply chain so as to provide quality products at competitive price in a timely manner. We are keeping up with the advanced techniques, growing through creating more values for our clients and society.
  • We always believe that the details decides the company's product quality, in this respect, the company conform our requirements and the goods are meet our expectations.
    5 Stars By Marcy Real from Peru - 2018.02.04 14:13
    The company has a good reputation in this industry, and finally it tured out that choose them is a good choice.
    5 Stars By Jenny from Norwegian - 2017.10.13 10:47
    Write your message here and send it to us