Functions |
ION(F-, CL-, Mg2+, Ca2+, NO3-, NH4+ etc) |
Measuring range |
0-20000ppm or 0-20ppm |
Resolution |
1ppm /0.01ppm |
Accuracy |
+/-1ppm, +/-0.01ppm |
mV input range |
0.00-1000.00mV |
Temp. compensation |
Pt 1000/NTC10K |
Temp range |
-10.0 to +130.0℃ |
Temp. Compensation range |
-10.0 to +130.0℃ |
Temp resolution |
0.1℃ |
Temp. accuracy |
±0.2℃ |
Ambient temperature range |
0 to +70℃ |
Storage temp |
-20 to +70℃ |
Input impedance |
>1012 Ω |
Display |
Back light, dot matrix |
ION current output1 |
Isolate, 4 to 20mA output, max load 500Ω |
Temp. current output 2 |
Isolate, 4 to 20mA output, max load 500Ω |
Current output accuracy |
±0.05 mA |
RS485 |
Modbus RTU protocol |
Baud rate |
9600/19200/38400 |
MAX. relay contacts capacity |
5A/250VAC , 5A/30VDC |
Cleaning setting |
On: 1 to 1000 seconds, Off: 0.1 to 1000.0 hours |
One multi-function relay |
clean/period alarm/error alarm |
Relay delay |
0-120 seconds |
Data logging capacity |
500,000 data |
Language selection |
English/traditional Chinese/simplified Chinese |
USB port |
Download records and update program |
IP Rating |
IP65 |
Power supply |
From 90 to 260 VAC, power consumption < 5 watts |
Installation |
panel/wall/pipe installation |
Weight |
0.85Kg |
An ion is a charged atom or molecule. It is charged because the number of electrons do not equal the number of protons in the atom or molecule. An atom can acquire a positive charge or a negative charge depending on whether the number of electrons in an atom is greater or less then the number of protons in the atom.
When an atom is attracted to another atom because it has an unequal number of electrons and protons, the atom is called an ION. If the atom has more electrons than protons, it is a negative ion, or ANION. If it has more protons than electrons,it is a positive ion.