Introduction
The built-in sensor has the characteristics of high measurement accuracy, fast response time and low maintenance cost. Standard 7-inch touch screen, the analyzer
outputs one 4-20mA standard signal and one RS485 signal. German Weidmuller terminals are used to ensure stable electrical connection. This product is easy to
install, high precision and small in size.
This product is widely used in industries where agricultural drinking water and water plants continuously monitor the residual chlorine content in aqueous solutions.
Technical Indexes
1. Display | 7" touch screen |
2. Measuring range | Residual chlorine: 0~5 mg/L;CLO2: 0-5mg/L |
3.Temperature | 0.1~40.0℃ |
4. Accuracy | ±2 %FS |
5. Response time | <30s |
6. Repeatability | ±0.02mg/L |
7. PH value range | 5~9pH |
8. Minimum conductivity | 100us/cm |
9. Water sample flow | 12~30L/H, in the flow cell |
10. Maximum pressure | 4bar |
11. Operating temperature | 0.1 to 40°C (without freezing) |
12. Output signal | 4-20mA |
13. Digital communication | equipped with MODBUS RS485 communication function, which can transmit measured values in real time |
14. Load resistance | ≤750Ω |
15. Ambient humidity | ≤95% no condensation |
16. Power supply | 220V AC |
17. Dimensions | 400×300×200mm |
18. Protection class | IP54 |
19. Window size | 155×87mm |