LSGG-5090Pro type industrial online phosphate analyzer, dopts special air rabbling and optoelectronics examination technique,

make chemical react quickly and measure accuracy outstanding, optoelectronics examination and chart text display. Adopt colorful

liquid crystal display, with the prolific color, character, chart and curve etc.

Can be widely used in thermal power plants, chemical industry and other departments, timely and accurate phosphate content of the

water is monitoring to ensure crew working safely, economic, especially for the scene environment.

 

Features:

1. 1 ~ 6 channels to for optional, cost savings.

2. High accuracy, fast response.

3. Regular automatic calibration, maintenance workload is small.

4. Color LCD real-time curve, convenient for analysis working condition.

5. Save a month of historical data, easy recall.

6. Monochromatic cold light source, long life, good stability.

7. Multiple-precision programmable current output, suitable for subsequent

automatic dosing or data acquisition system.

 

Technical Indexes

1. Measuring principle phosphorus molybdenum alum yellow photoelectric colorimetry
2. Measuring range 0~2000μg/L, 0~10mg/L (optional)
3. Accuracy ± 1% F.S
4. Reproducibility ± 1% F.S
5. Stability drift ≤ ± 1% F.S/24 hour
6. Response time  initial response, four minutes, six minutes to reach at least 98%
7. Sampling period 3 minutes/Channel
8. Watery conditions Flow> 2 ml / sec, Temperature: 10 ~ 45 ℃, Pressure: 10kPa ~ 100kPa
9. Ambient temperature 5 ~ 45 ℃ (higher than 40 ℃, reduced accuracy)
10.Environment humidity <85% RH
11. Reagent types one kind
12. Reagent consumption about 3 liters/month
13. Output signal 4-20mA
14. Alarm buzzer, relay normally open contacts
15.Communication RS-485、LAN、WIFI or 4G etc
16. Power supply AC220V±10%  50HZ
17. Power ≈50VA
18. Dimensions 720mm (height) × 460mm (width) × 300mm (depth)