1. Reliable, exact and fully automatic analysing
2. Simple commissioning with configuration assistant
3. Self calibrating and self monitoring
4. High measuring accuracy
5. Easy maintenance and cleaning.
6. Minimal reagent and water consumption
7. Multi-coloured and multi-lingual graphic display.
8. 0/4-20mA/relay/CAN-interface output
The Water Hardness/Alkali Analyzer are used in industrial measuring of water hardness and Alkali, such as Waste water treatment, environmental monitoring, drinking water and etc..
Hardness Reagents & Measurement Ranges
Reagent type | °dH | °F | ppm CaCO3 | mmol/l |
TH5001 | 0.03-0.3 | 0.053-0.534 | 0.534-5.340 | 0.005-0.053 |
TH5003 | 0.09-0.9 | 0.160-1.602 | 1.602-16.02 | 0.016-0.160 |
TH5010 | 0.3-3.0 | 0.534-5.340 | 5.340-53.40 | 0.053-0.535 |
TH5030 | 0.9-9.0 | 1.602-16.02 | 16.02-160.2 | 0.160-1.602 |
TH5050 | 1.5-15 | 2.67-26.7 | 26.7-267.0 | 0.267-2.670 |
TH5100 | 3.0-30 | 5.340-53.40 | 53.40-534.0 | 0.535-5.340 |
Alkali Reagents & Measurement Ranges
Reagents model | Measuring range |
TC5010 | 5.34~134 ppm |
TC5015 | 8.01~205ppm |
TC5020 | 10.7~267ppm |
TC5030 | 16.0~401ppm |
Specifications
Measurement method | Titration method |
Water inlet in general | clear, colourless, free of solid particles, without gas bubbles |
Measurement range | Hardness : 0.5-534ppm , total alkali:5.34~401ppm |
Accuracy | +/- 5% |
Repetition | ±2.5% |
Environmental temp. | 5-45℃ |
Measuring water temp. | 5-45℃ |
Water inlet pressure | ca. 0.5 - 5 bar (max.) (Recommended 1 - 2 bar) |
Analysis start | - programmable time intervals (5 - 360 minutes)- external signal
- programmable volume intervals |
Flush time | programmable flush time (15 - 1800 seconds) |
Output | - 4 x potential-free Relays (max. 250 Vac / Vdc; 4A(as potential free output NC/NO))- 0/4-20mA
- CAN interface |
Power | 90 - 260 Vac (47 - 63Hz) |
Power consumption | 25 VA (in operation), 3.5 VA (stand by) |
Dimensions | 300x300x200 mm (WxHxD) |
Protection grade | IP65 |
Write your message here and send it to us